7.35 ConductivitySignalType

The ConductivitySignalType provides ObjectType to add conductivity measurement specific variables. It is formally defined in Table 184.

Table 184 – ConductivitySignalType definition
Attribute Value
BrowseNameConductivitySignalType
IsAbstractFalse
References NodeClass BrowseName DataType TypeDefinition Other
Subtype of AnalyticalSignalType defined in 7.27, i.e. inheriting the InstanceDeclarations of that Node
0:HasComponentObject<SignalCalibrationIdentifier>0:BaseObjectTypeOP
0:HasComponentObjectSignalConditionSet0:BaseObjectTypeO
Conformance Units
PA-DIM ConductivitySignalType
PA-DIM ConductivitySignalType SensorCleaningsCounter
PA-DIM ConductivitySignalType SensorSterilisationsCounter

<SignalCalibrationIdentifier> is a container object for the calibration parameters of the signal. It is inherited from AnalogSignalType and has an additional interface reference to IConductivityCalibrationType to add conductivity signal specific calibration variables. Since it is a placeholder, an instance of ConductivitySignalType can have several calibration objects.

SignalConditionSet is a container object for the condition parameters of the signal. It is inherited from AnalogSignalType and has an additional interface reference to IConductivitySignalConditionSetType to add conductivity signal specific condition variables.

The components of ConductivitySignalType have additional references which are defined in Table 185.

Table 185 – ConductivitySignalType additional References
SourceBrowsePath Reference Type Is Forward TargetBrowsePath
<SignalCalibrationIdentifier>0:HasInterfaceTrueIConductivityCalibrationType
0:HasDictionaryEntryTrue 3:0112/2///61987#ABA968#004
SignalConditionSet0:HasInterfaceTrueIConductivitySignalConditionSetType
0:HasDictionaryEntryTrue 3:0112/2///61987#ABA968#004

The components of ConductivitySignalType have additional subcomponents which are defined in Table 186.

Table 186 – ConductivitySignalType additional subcomponents
BrowsePath References NodeClass BrowseName DataType TypeDefinition Others
Applied from IConductivityCalibrationType
<SignalCalibrationIdentifier>0:HasComponentVariableConductivityCellConstant0:Float0:AnalogUnitTypeO
Applied from IConductivitySignalConditionSetType
SignalConditionSet0:HasComponentVariableConductivityMeasuringMethod0:UInt320:MultiStateDictionaryEntryDiscreteTypeO
SignalConditionSet0:HasComponentVariableSensingElementTemperature0:Float0:AnalogUnitTypeO
SignalConditionSet0:HasComponentVariableTemperatureCompensationStyle0:UInt320:MultiStateDictionaryEntryDiscreteTypeO
SignalConditionSet0:HasPropertyVariableSensorCleaningsCounter0:UInt320:PropertyTypeO
SignalConditionSet0:HasPropertyVariableSensorSterilisationsCounter0:UInt320:PropertyTypeO

The child Nodes of the ConductivitySignalType have additional Attribute values defined in Table 187.

Table 187 – ConductivitySignalType Attribute values for child Nodes
BrowsePath Value Attribute
0.0
NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720
UnitId: 705744658
DisplayName: cm⁻¹
Description: reciprocal centimetre
0
ns=3;s=0112/2///61987#ABP721#001
inductive
0.0
NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720
UnitId: 705741427
DisplayName: °C
Description: degree Celsius
0
ns=3;s=0112/2///61987#ABP724#001
no temperature compensation
0
0