7.1.22 IPhSignalConditionSetType

The IPhSignalConditionSetType provides the interface to pH measurement signal specific condition variables and is formally defined in Table 76.

Table 76 - IPhSignalConditionSetType definition
AttributeValue
BrowseNameIPhSignalConditionSetType
IsAbstractTrue
ReferencesNodeClassBrowseNameDataTypeTypeDefinitionOther
Subtype of 0:BaseInterfaceType defined in OPC 10000-5, i.e. inheriting the InstanceDeclarations of that Node
0:HasComponentVariablePhMeasuringMethod0:UInt320:MultiStateDictionaryEntryDiscreteTypeO
0:HasComponentVariableSensingElementTemperature0:Float0:AnalogUnitTypeO
0:HasComponentVariableSensorNextCalibration0:UInt320:AnalogUnitTypeO
0:HasComponentVariableSensingElementImpedance0:Float0:AnalogUnitTypeO
0:HasComponentVariableSensorReferenceImpedance0:Float0:AnalogUnitTypeO
0:HasPropertyVariableSensorCleaningsCounter0:UInt320:PropertyTypeO
0:HasPropertyVariableSensorSterilisationsCounter0:UInt320:PropertyTypeO
Conformance Units
PA-DIM PhSignalType
PA-DIM PhSignalType SensorNextCalibration
PA-DIM PhSignalType SensingElementImpedance
PA-DIM PhSignalType SensorReferenceImpedance
PA-DIM PhSignalType SensorCleaningsCounter
PA-DIM PhSignalType SensorSterilisationsCounter

PhMeasuringMethod is defined by IRDI as ABP640#002 which states "physical measuring method of a pH sensor".

SensingElementTemperature is defined by IRDI as ABP565#001 which states "temperature measured at a sensing element".

SensorNextCalibration is defined by IRDI as ABP566#001 which states "time that can elapse until the next calibration, calculated from the configured calibration interval or estimated by the device".

SensingElementImpedance is defined by IRDI as ABP570#001 which states "absolute value of impedance for a pH sensing element, compensated for a temperature of 25°C".

SensorReferenceImpedance is defined by IRDI as ABP571#001 which states "absolute value of impedance for a pH reference system, compensated for a temperature of 25°C".

SensorCleaningsCounter is defined by IRDI as ABP546#001 which states "number of times a sensor has been cleaned in place".

SensorSterilisationsCounter is defined by IRDI as ABP547#001 which states "number of times a sensor has been sterilized in place".

The components of IPhSignalConditionSetType have additional references which are defined in Table 77.

Table 77 - IPhSignalConditionSetType additional References
SourceBrowsePathReference TypeIs ForwardTargetBrowsePath
PhMeasuringMethod0:HasDictionaryEntryTrue3:0112/2///61987#ABP640#002
SensingElementTemperature0:HasDictionaryEntryTrue3:0112/2///61987#ABP565#001
SensingElementTemperature
0:EngineeringUnits

0:HasDictionaryEntryTrue3:0112/2///61987#ABA968#004
SensorNextCalibration0:HasDictionaryEntryTrue3:0112/2///61987#ABP566#001
SensorNextCalibration
0:EngineeringUnits

0:HasDictionaryEntryTrue3:0112/2///61987#ABA968#004
SensingElementImpedance0:HasDictionaryEntryTrue3:0112/2///61987#ABP570#001
SensingElementImpedance
0:EngineeringUnits

0:HasDictionaryEntryTrue3:0112/2///61987#ABA968#004
SensorReferenceImpedance0:HasDictionaryEntryTrue3:0112/2///61987#ABP571#001
SensorReferenceImpedance
0:EngineeringUnits

0:HasDictionaryEntryTrue3:0112/2///61987#ABA968#004
SensorCleaningsCounter0:HasDictionaryEntryTrue3:0112/2///61987#ABP546#001
SensorSterilisationsCounter0:HasDictionaryEntryTrue3:0112/2///61987#ABP547#001

The child Nodes of the IPhSignalConditionSetType have additional Attribute values defined in Table 78.

Table 78 - IPhSignalConditionSetType Attribute values for child Nodes
BrowsePathValue Attribute
PhMeasuringMethod0
PhMeasuringMethod
0:ValueAsDictionaryEntries

ns=3;s=0112/2///61987#ABP718#001
PhMeasuringMethod
0:ValueAsText

glass electrode
SensingElementTemperature0.0
SensingElementTemperature
0:EngineeringUnits

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720
UnitId: 705741427
DisplayName: °C

Description: degree Celsius

SensorNextCalibration0
SensorNextCalibration
0:EngineeringUnits

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720

UnitId: 705745431

DisplayName: d

Description: day

SensingElementImpedance0.0
SensingElementImpedance
0:EngineeringUnits

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720

UnitId: 705741367

DisplayName: Ω

Description: ohm

SensorReferenceImpedance0.0
SensorReferenceImpedance
0:EngineeringUnits

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720

UnitId: 705741367

DisplayName: Ω

Description: ohm

SensorCleaningsCounter0
SensorSterilisationsCounter0