7.1.12 IGasChromatographDeviceConditionSetType

The IGasChromatographDeviceConditionSetType provides the interface to gas chromatograph device specific condition variables and is formally defined in Table 46

Table 46 – IGasChromatographDeviceConditionSetType definition
Attribute Value
BrowseNameIGasChromatographDeviceConditionSetType
IsAbstractTrue
References NodeClass BrowseName DataType TypeDefinition Other
Subtype of 0:BaseInterfaceType defined in OPC 10000-5, i.e. inheriting the InstanceDeclarations of that Node
0:HasPropertyVariableValveName0:LocalizedText[]0:PropertyTypeO
0:HasPropertyVariableValveSwitchingCyclesCounter0:UInt32[]0:PropertyTypeO
0:HasComponentVariableTotalAreaMeasuredPeaks0:Float0:AnalogUnitTypeO
0:HasComponentVariableBaselineNoise0:Float0:AnalogUnitTypeO
Conformance Units
PA-DIM GasChromatographType
PA-DIM GasChromatographType Valve
PA-DIM GasChromatographType TotalAreaMeasuredPeaks
PA-DIM GasChromatographType BaselineNoise

The components of IGasChromatographDeviceConditionSetType have additional references which are defined in Table 48.

Table 47 – IGasChromatographDeviceConditionSetType additional References
SourceBrowsePath Reference Type Is Forward TargetBrowsePath
ValveName0:HasDictionaryEntryTrue 3:0112/2///61987#ABQ046#001
ValveSwitchingCyclesCounter0:HasDictionaryEntryTrue 3:0112/2///61987#ABQ007#001
TotalAreaMeasuredPeaks0:HasDictionaryEntryTrue 3:0112/2///61987#ABQ043#001
BaselineNoise0:HasDictionaryEntryTrue 3:0112/2///61987#ABQ036#001

The child Nodes of the IGasChromatographDeviceConditionSetType have additional Attribute values defined in Table 48.

Table 48 – IGasChromatographDeviceConditionSetType Attribute values for child Nodes
BrowsePath Value Attribute
TotalAreaMeasuredPeaks0.0

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720

UnitId: 705741328

DisplayName: %

Description: percent

BaselineNoise0.0

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720
UnitId: 705749524

DisplayName: mV
Description: millivolt