7.1.17 ITocSignalConditionSetType

The ITocSignalConditionSetType provides the interface to TOC measurement signal specific condition variables and is formally defined in Table 61.

Table 61 – ITocSignalConditionSetType definition
Attribute Value
BrowseNameITocSignalConditionSetType
IsAbstractTrue
References NodeClass BrowseName DataType TypeDefinition Other
Subtype of 0:BaseInterfaceType defined in OPC 10000-5, i.e. inheriting the InstanceDeclarations of that Node
0:HasComponentVariableChopperFrequencyDeviation0:Float0:BaseAnalogTypeO
0:HasComponentVariableRelativeReagentLevel0:Float{Any}0:AnalogUnitTypeO
0:HasComponentVariableSampleCellTemperature0:Float0:AnalogUnitTypeO
0:HasComponentVariableDetectorZeroSignal0:Float0:AnalogUnitTypeO
0:HasPropertyVariableSourceResidualLife0:Float0:PropertyTypeO
0:HasComponentVariableAbsoluteSampleGasPressure0:Float0:AnalogUnitTypeO
0:HasComponentVariableSampleGasVolumeFlow0:Float0:AnalogUnitTypeO
Conformance Units
PA-DIM TocSignalType
PA-DIM TocSignalType ChopperFrequencyDeviation
PA-DIM TocSignalType RelativeReagentLevel
PA-DIM TocSignalType AbsoluteSampleGasPressure
PA-DIM TocSignalType SampleGasVolumeFlow

The components of ITocSignalConditionSetType have additional references which are defined in Table 62.

Table 62 – ITocSignalConditionSetType additional References
SourceBrowsePath Reference Type Is Forward TargetBrowsePath
ChopperFrequencyDeviation0:HasDictionaryEntryTrue 3:0112/2///61987#ABP553#001
0:HasDictionaryEntryTrue 3:0112/2///61987#ABA968#004
RelativeReagentLevel0:HasDictionaryEntryTrue 3:0112/2///61987#ABP557#001
0:HasDictionaryEntryTrue 3:0112/2///61987#ABA968#004
SampleCellTemperature0:HasDictionaryEntryTrue 3:0112/2///61987#ABP556#001
0:HasDictionaryEntryTrue 3:0112/2///61987#ABA968#004
DetectorZeroSignal0:HasDictionaryEntryTrue 3:0112/2///61987#ABP551#001
0:HasDictionaryEntryTrue 3:0112/2///61987#ABA968#004
SourceResidualLife0:HasDictionaryEntryTrue 3:0112/2///61987#ABP552#001
AbsoluteSampleGasPressure0:HasDictionaryEntryTrue 3:0112/2///61987#ABP560#001
0:HasDictionaryEntryTrue 3:0112/2///61987#ABA968#004
SampleGasVolumeFlow0:HasDictionaryEntryTrue 3:0112/2///61987#ABP562#001
0:HasDictionaryEntryTrue 3:0112/2///61987#ABA968#004

The child Nodes of the ITocSignalConditionSetType have additional Attribute values defined in Table 63.

Table 63 – ITocSignalConditionSetType Attribute values for child Nodes
BrowsePath Value Attribute
ChopperFrequencyDeviation0.0

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720

UnitId: 705741328

DisplayName: %

Description: percent

RelativeReagentLevel0.0

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720

UnitId: 705741328

DisplayName: %

Description: percent

SampleCellTemperature0.0
NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720
UnitId: 705741427
DisplayName: °C
Description: degree Celsius
DetectorZeroSignal0.0

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720
UnitId: 705848917

DisplayName: ppm

Description: parts per million

SourceResidualLife1.0
AbsoluteSampleGasPressure0.0
NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720
UnitId: 705749552
DisplayName: mbar
Description: millibar
SampleGasVolumeFlow0.0

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720

UnitId: 705747637

DisplayName: l/h

Description: litre per hour