10.13 CalibrationDataType
This structure contains the Calibration information. It is used as an input argument in SetCalibration method.
Note: The input data sent in SetCalibration shall be updated in the respective parameters of the asset under Maintenance/Calibration.
The structure is defined in Table 232.
| Name | Type | Description | Optional |
| CalibrationDataType | structure | Subtype of the 0:Structure defined in OPC 10000-5. | |
| LastCalibration | 0:UtcTime | It is the date when the last calibration took place. Note: It can have the initial operation date for a new asset if it is not available. | False |
| CalibrationPlace | 0:String | It is the location where the last calibration took place. | True |
NextCalibration | 0:UtcTime | It is the date of the next planned calibration. | True |
| CalibrationValue | 0:Double | It is the actual measured value of the calibration. | True |
| SensorScale | 0:Double | It is the nominal scale of the sensor. It corresponds also with the measurement range of the sensor. | True |
| CertificateUri | 0:UriString | It contains the URI of a certificate of the calibration target in case the calibration target is certified and the information available. Otherwise, the Variable should be omitted. The String shall be a URI as defined by RFC 3986. Example: MCE test document. | True |
| EngineeringUnits | 0:EUInformation | It defines the engineering unit of the value. | True |
Its representation in the AddressSpace is defined in Table 233.
| Attribute | Value | ||||
| BrowseName | CalibrationDataType | ||||
| IsAbstract | False | ||||
| References | Node Class | BrowseName | DataType | TypeDefinition | Other |
|---|---|---|---|---|---|
| Subtype of the 0:Structure defined in OPC 10000-5. | |||||
| Conformance Units | |||||
|---|---|---|---|---|---|
| IJT Joining System Base | |||||
| IJT Engineering Units |