The IPhSignalConditionSetType provides the interface to pH measurement signal specific condition variables and is formally defined in Table 76.

Table 76 – IPhSignalConditionSetType definition

Attribute

Value

BrowseName

IPhSignalConditionSetType

IsAbstract

True

References

NodeClass

BrowseName

DataType

TypeDefinition

Other

Subtype of 0:BaseInterfaceType defined in OPC 10000-5, i.e. inheriting the InstanceDeclarations of that Node

0:HasComponent

Variable

PhMeasuringMethod

0:UInt32

0:MultiStateDictionaryEntryDiscreteType

O

0:HasComponent

Variable

SensingElementTemperature

0:Float

0:AnalogUnitType

O

0:HasComponent

Variable

SensorNextCalibration

0:UInt32

0:AnalogUnitType

O

0:HasComponent

Variable

SensingElementImpedance

0:Float

0:AnalogUnitType

O

0:HasComponent

Variable

SensorReferenceImpedance

0:Float

0:AnalogUnitType

O

0:HasProperty

Variable

SensorCleaningsCounter

0:UInt32

0:PropertyType

O

0:HasProperty

Variable

SensorSterilisationsCounter

0:UInt32

0:PropertyType

O

Conformance Units

PA-DIM PhSignalType

PA-DIM PhSignalType SensorNextCalibration

PA-DIM PhSignalType SensingElementImpedance

PA-DIM PhSignalType SensorReferenceImpedance

PA-DIM PhSignalType SensorCleaningsCounter

PA-DIM PhSignalType SensorSterilisationsCounter

PhMeasuringMethod is defined by IRDI as ABP640#002 which states “physical measuring method of a pH sensor”.

SensingElementTemperature is defined by IRDI as ABP565#001 which states “temperature measured at a sensing element”.

SensorNextCalibration is defined by IRDI as ABP566#001 which states “time that can elapse until the next calibration, calculated from the configured calibration interval or estimated by the device”.

SensingElementImpedance is defined by IRDI as ABP570#001 which states “absolute value of impedance for a pH sensing element, compensated for a temperature of 25°C”.

SensorReferenceImpedance is defined by IRDI as ABP571#001 which states “absolute value of impedance for a pH reference system, compensated for a temperature of 25°C”.

SensorCleaningsCounter is defined by IRDI as ABP546#001 which states “number of times a sensor has been cleaned in place”.

SensorSterilisationsCounter is defined by IRDI as ABP547#001 which states “number of times a sensor has been sterilized in place”.

The components of IPhSignalConditionSetType have additional references which are defined in Table 77.

Table 77 – IPhSignalConditionSetType additional References

SourceBrowsePath

Reference Type

Is Forward

TargetBrowsePath

PhMeasuringMethod

0:HasDictionaryEntry

True

3:0112/2///61987#ABP640#002

SensingElementTemperature

0:HasDictionaryEntry

True

3:0112/2///61987#ABP565#001

SensingElementTemperature

0:EngineeringUnits

0:HasDictionaryEntry

True

3:0112/2///61987#ABA968#004

SensorNextCalibration

0:HasDictionaryEntry

True

3:0112/2///61987#ABP566#001

SensorNextCalibration

0:EngineeringUnits

0:HasDictionaryEntry

True

3:0112/2///61987#ABA968#004

SensingElementImpedance

0:HasDictionaryEntry

True

3:0112/2///61987#ABP570#001

SensingElementImpedance

0:EngineeringUnits

0:HasDictionaryEntry

True

3:0112/2///61987#ABA968#004

SensorReferenceImpedance

0:HasDictionaryEntry

True

3:0112/2///61987#ABP571#001

SensorReferenceImpedance

0:EngineeringUnits

0:HasDictionaryEntry

True

3:0112/2///61987#ABA968#004

SensorCleaningsCounter

0:HasDictionaryEntry

True

3:0112/2///61987#ABP546#001

SensorSterilisationsCounter

0:HasDictionaryEntry

True

3:0112/2///61987#ABP547#001

The child Nodes of the IPhSignalConditionSetType have additional Attribute values defined in Table 78.

Table 78 – IPhSignalConditionSetType Attribute values for child Nodes

BrowsePath

Value Attribute

PhMeasuringMethod

0

PhMeasuringMethod

0:ValueAsDictionaryEntries

ns=3;s=0112/2///61987#ABP718#001

PhMeasuringMethod

0:ValueAsText

glass electrode

SensingElementTemperature

0.0

SensingElementTemperature

0:EngineeringUnits

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720UnitId: 705741427 DisplayName: °C

Description: degree Celsius

SensorNextCalibration

0

SensorNextCalibration

0:EngineeringUnits

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720

UnitId: 705745431

DisplayName: d

Description: day

SensingElementImpedance

0.0

SensingElementImpedance

0:EngineeringUnits

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720

UnitId: 705741367

DisplayName: Ω

Description: ohm

SensorReferenceImpedance

0.0

SensorReferenceImpedance

0:EngineeringUnits

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720

UnitId: 705741367

DisplayName: Ω

Description: ohm

SensorCleaningsCounter

0

SensorSterilisationsCounter

0