Search
3 result(s) for RawValue
-
OPC-30500-1 – OPC UA for Laboratory & Analytical Device Standard (LADS) - Part 1: BasicsBaseSensorFunctionType defined in 7.5.1 0:HasComponent Object AlarmMonitor ExclusiveLevelAlarmType O 0:HasComponent Variable RawValue Double{Any} AnalogUnitRangeType M 0:HasComponent Variable SensorValue Double{Any} AnalogUnitRangeType M 0:HasComponent Object Calibration ... limit of an analogue Sensor is exceeded. See: 10000-9: Alarms & Conditions | ExclusiveLevelAlarmType . RawValue is the raw value measured at the Sensor element, such as the Nernst voltage
-
OPC-30500-1 – OPC UA for Laboratory & Analytical Device Standard (LADS) - Part 1: BasicsBrowseName DataType TypeDefinition Other Subtype of the AnalogSensorFunctionType defined in 7.5.2 0:HasComponent Variable RawValue Double{OneOrMoreDimensions} AnalogUnitRangeType M 0:HasComponent Variable SensorValue Double{OneOrMoreDimensions} AnalogUnitRangeType M Conformance Units LADS ... AnalogArraySensorFunctionType RawValue is the raw value measured at the Sensor element, such as the Nernst voltage of a pH Sensor element. SensorValue is the calibrated and optionally compensated/filtered process value
-
OPC-30500-1 – OPC UA for Laboratory & Analytical Device Standard (LADS) - Part 1: BasicsBrowseName DataType TypeDefinition Other Subtype of the AnalogSensorFunctionType defined in 7.5.2 0:HasComponent Variable RawValue Double AnalogUnitRangeType M 0:HasComponent Variable SensorValue Double AnalogUnitRangeType M Conformance Units LADS AnalogScalarSensorFunctionType RawValue