The IAmperometricSignalConditionSetType provides the interface to amperometric measurement signal specific condition variables and is formally defined in Table 67.
Table 67 – IAmperometricSignalConditionSetType definition
Attribute |
Value |
||||
BrowseName |
IAmperometricSignalConditionSetType |
||||
IsAbstract |
True |
||||
References |
NodeClass |
BrowseName |
DataType |
TypeDefinition |
Other |
Subtype of 0:BaseInterfaceType defined in OPC 10000-5, i.e. inheriting the InstanceDeclarations of that Node |
|||||
0:HasComponent |
Variable |
SensingElementTemperature |
0:Float |
0:AnalogUnitType |
O |
0:HasComponent |
Variable |
SensorNextCalibration |
0:UInt32 |
0:AnalogUnitType |
O |
0:HasProperty |
Variable |
SensorCleaningsCounter |
0:UInt32 |
0:PropertyType |
O |
0:HasProperty |
Variable |
SensorSterilisationsCounter |
0:UInt32 |
0:PropertyType |
O |
Conformance Units |
|||||
PA-DIM AmperometricSignalType |
|||||
PA-DIM AmperometricSignalType SensorNextCalibration |
|||||
PA-DIM AmperometricSignalType SensorCleaningsCounter |
|||||
PA-DIM AmperometricSignalType SensorSterilisationsCounter |
SensingElementTemperature is defined by IRDI as ABP565#001 which states “temperature measured at a sensing element”.
SensorNextCalibration is defined by IRDI as ABP566#001 which states “time that can elapse until the next calibration, calculated from the configured calibration interval or estimated by the device”.
SensorCleaningsCounter is defined by IRDI as ABP546#001 which states “number of times a sensor has been cleaned in place”.
SensorSterilisationsCounter is defined by IRDI as ABP547#001 which states “number of times a sensor has been sterilized in place”.
The components of IAmperometricSignalConditionSetType have additional references which are defined in Table 68.
Table 68 – IAmperometricSignalConditionSetType additional References
SourceBrowsePath |
Reference Type |
Is Forward |
TargetBrowsePath |
||
SensingElementTemperature |
0:HasDictionaryEntry |
True |
3:0112/2///61987#ABP565#001 |
||
|
0:HasDictionaryEntry |
True |
|||
SensorNextCalibration |
0:HasDictionaryEntry |
True |
3:0112/2///61987#ABP566#001 |
||
|
0:HasDictionaryEntry |
True |
|||
SensorCleaningsCounter |
0:HasDictionaryEntry |
True |
3:0112/2///61987#ABP546#001 |
||
SensorSterilisationsCounter |
0:HasDictionaryEntry |
True |
3:0112/2///61987#ABP547#001 |
The child Nodes of the IAmperometricSignalConditionSetType have additional Attribute values defined in Table 69.
Table 69 – IAmperometricSignalConditionSetType Attribute values for child Nodes
BrowsePath |
Value Attribute |
||
SensingElementTemperature |
0.0 |
||
|
NamespaceUri: http://www.opcfoundation.org/UA/units/un/cefactUnitId: 4408652DisplayName: °CDescription: degree Celsius |
||
SensorNextCalibration |
0 |
||
|
NamespaceUri: http://www.opcfoundation.org/UA/units/un/cefactUnitId: 4473177DisplayName: dDescription: day |
||
SensorCleaningsCounter |
0 |
||
SensorSterilisationsCounter |
0 |