The ModifyTargetValueBy Method is used to modify the current TargetValue by a relative amount. The resulting value shall be limited to the TargetValue’s allowed range (EURange).

The direction of the action is application dependent and is controlled by the sign of the provided value. For instance, negative values may indicate moving left and positive values may indicate moving right, or positive values may be used for aspirating a defined volume while negative values may be used for dispensing a defined volume.

The signature of this Method is specified below. Table 101 and Table 102 specify the Arguments and AddressSpace representation, respectively.

Signature

ModifyTargetValueBy (

[in]0:DoubleValue

)

Table 101 – ModifyTargetValueBy Method Arguments

Argument

Description

Value

Relative value by which the target value will be changed. The resulting value will typically be limited to the target-value’s allowed range. Provided values can be positive or negative.

Table 102 – ModifyTargetValueBy Method AddressSpace Definition

Attribute

Value

BrowseName

ModifyTargetValueBy

References

Node Class

BrowseName

DataType

TypeDefinition

ModellingRule

0:HasProperty

Variable

0:InputArguments

0:Argument[]

0:PropertyType

0:Mandatory