Annex A (informative) – Example of extending ADI Information Model for particle size monitor devices ToC Previous Next

A.2 Parameters of ParticleSizeMonitorDeviceType ToC Previous Next

A.2.2 AnalyserChannel of ParticleSizeMonitorDeviceType (General Approach) ToC Previous Next

As an alternative to chapter A.2.1, the tables below show a more general approach of defining the Parameters, independent of the used technology.

Table 96 – ParticleSizeMonitorDeviceType AnalyserChannel Status Parameters (Alternative to Table 94)

BrowseName Description VariableType RW Optional/
Mandatory
InstrumentConnected Return the status of the physical connection with the channel TwoStateDiscreteType
(DataType=Boolean)
RO O
ReadyForBackground Return the status of the instrument and accessories to perform a background reading TwoStateDiscreteType
(DataType=Boolean)
RO O
ReadyForMeasurement Return the status of the instrument and accessories to perform a measurement TwoStateDiscreteType
(DataType=Boolean)
RO O

All Parameters organized by the Status FunctionalGroup on an AnalyserChannel of a ParticleSizeMonitorDeviceType shall be read-only.

Table 97 – ParticleSizeMonitorDeviceStreamType AcquisitionSettings Parameters (Alternative to Table 95)

BrowseName Description VariableType Optional/
Mandatory
AcquisitionSettings Name of a set of acquisition settings stored in the analyser. String M

Previous Next