The IGasChromatographCalibrationType provides the interface to gaschromatograph measurement specific calibration variables of the sensor resp. signal and is formally defined in Table 32.

Table 32 – IGasChromatographCalibrationTypedefinition

Attribute

Value

BrowseName

IGasChromatographCalibrationType

IsAbstract

True

References

NodeClass

BrowseName

DataType

TypeDefinition

Other

Subtype of 0:BaseInterfaceType defined in OPC 10000-5, i.e. inheriting the InstanceDeclarations of that Node

0:HasProperty

Variable

CalibrationRange1ResponseFactor

0:Float

0:PropertyType

O

0:HasComponent

Variable

CalibrationRange1LowerRangeValue

0:Float

0:AnalogUnitType

O

0:HasComponent

Variable

CalibrationRange1UpperRangeValue

0:Float

0:AnalogUnitType

O

0:HasProperty

Variable

CalibrationRange2ResponseFactor

0:Float

0:PropertyType

O

0:HasComponent

Variable

CalibrationRange2LowerRangeValue

0:Float

0:AnalogUnitType

O

0:HasComponent

Variable

CalibrationRange2UpperRangeValue

0:Float

0:AnalogUnitType

O

0:HasProperty

Variable

CalibrationRange3ResponseFactor

0:Float

0:PropertyType

O

0:HasComponent

Variable

CalibrationRange3LowerRangeValue

0:Float

0:AnalogUnitType

O

0:HasComponent

Variable

CalibrationRange3UpperRangeValue

0:Float

0:AnalogUnitType

O

Conformance Units

PA-DIM GasChromatograph CalibrationRange1

PA-DIM GasChromatograph CalibrationRange2

PA-DIM GasChromatograph CalibrationRange3

CalibrationRange1ResponseFactor is defined by IRDI as ABQ024#001 which states "ratio between the concentration of a compound being analysed and the response of the detector to that compound for the calibration range 1".

CalibrationRange1LowerRangeValue is defined by IRDI as ABQ025#001 which states "volume concentration value assigned to the lower range end-value of calibration range 1".

CalibrationRange1UpperRangeValue is defined by IRDI as ABQ026#001 which states "volume concentration value assigned to the upper range end-value of calibration range 1".

CalibrationRange2ResponseFactor is defined by IRDI as ABQ027#001 which states "ratio between the concentration of a compound being analysed and the response of the detector to that compound for the calibration range 2".

CalibrationRange2LowerRangeValue is defined by IRDI as ABQ028#001 which states "volume concentration value assigned to the lower range end-value of calibration range 2".

CalibrationRange2UpperRangeValue is defined by IRDI as ABQ029#001 which states "volume concentration value assigned to the upper range end-value of calibration range 2".

CalibrationRange3ResponseFactor is defined by IRDI as ABQ030#001 which states "ratio between the concentration of a compound being analysed and the response of the detector to that compound for the calibration range 3".

CalibrationRange3LowerRangeValue is defined by IRDI as ABQ031#001 which states "volume concentration value assigned to the lower range end-value of calibration range 3".

CalibrationRange3UpperRangeValue is defined by IRDI as ABQ032#001 which states "volume concentration value assigned to the upper range end-value of calibration range 3".

The components of IGasChromatographCalibrationType have additional references which are defined in Table 33.

Table 33 – IGasChromatographCalibrationType additional References

SourceBrowsePath

Reference Type

Is Forward

TargetBrowsePath

CalibrationRange1ResponseFactor

0:HasDictionaryEntry

True

3:0112/2///61987#ABQ024#001

CalibrationRange1LowerRangeValue

0:HasDictionaryEntry

True

3:0112/2///61987#ABQ025#001

CalibrationRange1UpperRangeValue

0:HasDictionaryEntry

True

3:0112/2///61987#ABQ026#001

CalibrationRange2ResponseFactor

0:HasDictionaryEntry

True

3:0112/2///61987#ABQ027#001

CalibrationRange2LowerRangeValue

0:HasDictionaryEntry

True

3:0112/2///61987#ABQ028#001

CalibrationRange2UpperRangeValue

0:HasDictionaryEntry

True

3:0112/2///61987#ABQ029#001

CalibrationRange3ResponseFactor

0:HasDictionaryEntry

True

3:0112/2///61987#ABQ030#001

CalibrationRange3LowerRangeValue

0:HasDictionaryEntry

True

3:0112/2///61987#ABQ031#001

CalibrationRange3UpperRangeValue

0:HasDictionaryEntry

True

3:0112/2///61987#ABQ032#001

The child Nodes of the IGasChromatographCalibrationType have additional Attribute values defined in Table 34.

Table 34 – IGasChromatographCalibrationType Attribute values for child Nodes

BrowsePath

Value Attribute

CalibrationRange1ResponseFactor

1.0

CalibrationRange1LowerRangeValue

0.0

CalibrationRange1LowerRangeValue

0:EngineeringUnits

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720

UnitId: 705848917

DisplayName: ppm

Description: parts per million

CalibrationRange1UpperRangeValue

100.0

CalibrationRange1UpperRangeValue

0:EngineeringUnits

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720

UnitId: 705848917

DisplayName: ppm

Description: parts per million

CalibrationRange2ResponseFactor

1.0

CalibrationRange2LowerRangeValue

0.0

CalibrationRange2LowerRangeValue

0:EngineeringUnits

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720

UnitId: 705848917

DisplayName: ppm

Description: parts per million

CalibrationRange2UpperRangeValue

100.0

CalibrationRange2UpperRangeValue

0:EngineeringUnits

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720

UnitId: 705848917

DisplayName: ppm

Description: parts per million

CalibrationRange3ResponseFactor

1.0

CalibrationRange3LowerRangeValue

0.0

CalibrationRange3LowerRangeValue

0:EngineeringUnits

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720

UnitId: 705848917

DisplayName: ppm

Description: parts per million

CalibrationRange3UpperRangeValue

100.0

CalibrationRange3UpperRangeValue

0:EngineeringUnits

NamespaceUri: http://www.opcfoundation.org/UA/units/cdd/IEC62720

UnitId: 705848917

DisplayName: ppm

Description: parts per million