The concentration of a given Component and its related characteristics are represented using a Parameter of a type derived from EngineeringValueType.
The Value DataType is Float and its ValueRank is -1 because it is a scalar.
All components of this Variable are read-only.
Table 111 and Table 112 illustrate two example definitions of types that represent gas chromatograph Components.
Table 111 - ABBComponentValueType definition
Attribute |
Value |
|||||
BrowseName |
ABBComponentValueType |
|||||
IsAbstract |
False |
|||||
References |
NodeClass |
BrowseName |
Description |
TypeDefinition |
ModellingRule |
|
Subtype of the EngineeringValueType |
||||||
HasComponent |
Variable |
AmplitudeEOB |
Detector signal amplitude for the end of baseline calculation |
AnalogItemType (DataType=Float) |
Mandatory |
|
HasComponent |
Variable |
AmplitudeEOI |
Detector signal amplitude for the end of integration calculation |
AnalogItemType (DataType=Float) |
Mandatory |
|
HasComponent |
Variable |
AmplitudeSOB |
Detector signal amplitude for the start of baseline calculation |
AnalogItemType (DataType=Float) |
Mandatory |
|
HasComponent |
Variable |
AmplitudeSOI |
Detector signal amplitude for the start of integration calculation |
AnalogItemType (DataType=Float) |
Mandatory |
|
HasComponent |
Variable |
BaselineEnd |
Time into analysis of end of baseline calculation |
AnalogItemType (DataType=Float) |
Mandatory |
|
HasComponent |
Variable |
BaselineStart |
Time into analysis of start of baseline calculation |
AnalogItemType (DataType=Float) |
Mandatory |
|
HasComponent |
Variable |
BenchmarkDeviation |
This component’s deviation from defined validation concentration |
AnalogItemType (DataType=Float) |
Mandatory |
|
HasComponent |
Variable |
CrestAmplitude |
Maximum peak height for this component |
AnalogItemType (DataType=Float) |
Mandatory |
|
HasComponent |
Variable |
ExpectedConcentration |
This component’s expected concentration result from a validation analysis |
AnalogItemType (DataType=Float) |
Mandatory |
|
HasComponent |
Variable |
ExpectedRT |
This component’s expected retention time for a given analysis |
AnalogItemType (DataType=Float) |
Mandatory |
|
HasComponent |
Variable |
IntegrationEnd |
Time into analysis of end of integration calculation |
AnalogItemType (DataType=Float) |
Mandatory |
|
HasComponent |
Variable |
IntegrationStart |
Time into analysis of start of integration calculation |
AnalogItemType (DataType=Float) |
Mandatory |
|
HasComponent |
Variable |
IsValid |
Validity flag for this component |
TwoStateDiscreteType (DataType=Boolean) |
Mandatory |
|
HasComponent |
Variable |
NegativeArea |
Negative peak area for this component |
AnalogItemType (DataType=Float) |
Mandatory |
|
HasComponent |
Variable |
OldResponseFactor |
Previous calibration response factor for this component |
AnalogItemType (DataType=Float) |
Mandatory |
|
HasComponent |
Variable |
PeakFound |
Flag for a peak detected |
TwoStateDiscreteType (DataType=Boolean) |
Mandatory |
|
HasComponent |
Variable |
PositiveArea |
Positive peak area for this component |
AnalogItemType (DataType=Float) |
Mandatory |
|
HasComponent |
Variable |
ResponseFactor |
Calibration response factor for this component and associated detector |
AnalogItemType (DataType=Float) |
Mandatory |
|
HasComponent |
Variable |
Retention Time |
Actual retention time at the peak apex for this component |
AnalogItemType (DataType=Float) |
Mandatory |
|
HasComponent |
Variable |
RFUpdated |
Flag if the new RF from a calibration analysis is accepted |
TwoStateDiscreteType (DataType=Boolean) |
Mandatory |
|
HasComponent |
Variable |
TotalArea |
Total peak area for this component |
AnalogItemType (DataType=Float) |
Mandatory |
Table 112 – SiemensComponentValueType Definition
Attribute |
Value |
||||||||
BrowseName |
SiemensComponentValueType |
||||||||
IsAbstract |
False |
||||||||
References |
NodeClass |
BrowseName |
Description |
TypeDefinition |
ModellingRule |
||||
Subtype of the EngineeringValueType |
|||||||||
HasComponent |
Variable |
PkArea |
Corrected peak area for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkArea% |
Percent area for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkAsym |
Peak asymmetry for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkAsym10 |
Asymmetry at 10% peak height for |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkHeight |
Maximum peak height for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkHeight% |
Percent height for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkNoise |
Peak noise for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkResolution |
Peak resolution relative to previous peak for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkRetTime |
Retention time at peak apex for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkSignaltoNoise |
Peak signal to noise for this component. (Peak height / rms noise) |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkStartFlg |
Peak start flag for this component. |
AnalogItemType (DataType=String) |
Mandatory |
||||
HasComponent |
Variable |
PkStartTime |
Retention time at start of peak for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkStopFlg |
Peak stop flag for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkStopTime |
Retention time at end of peak for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkType |
Peak type this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkUSPWidth |
Peak USP width (U.S. Pharmacopeia) for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkWidth |
Peak width at base for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkWidth10 |
Peak width at 10% height for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkWidth5% |
Peak width at 5% height for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkWidth50 |
Peak width at 50% height for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
PkTheorPlates |
Theoretical plates for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
GrpArea |
Corrected peak area (for the group) for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
GrpArea% |
Peak area percent (for the group) for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
GrpHeight |
Maximum peak height (for the group) for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
RespFactL0 |
Calibration level response factor for level 0 if applicable for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
RespFactL1 |
Calibration level response factor for level 1 if applicable for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
RespFactL2 |
Calibration level response factor for level 2 if applicable for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
RespFactL3 |
Calibration level response factor for level 3 if applicable for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
RespFactL4 |
Calibration level response factor for level 4 if applicable for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
RespFactL5 |
Calibration level response factor for level 5 if applicable for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
RespFactL6 |
Calibration level response factor for level 6 if applicable for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
RespFactL7 |
Calibration level response factor for level 7 if applicable for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
RespFactL8 |
Calibration level response factor for level 8 if applicable for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
||||
HasComponent |
Variable |
RespFactL9 |
Calibration level response factor for level 9 if applicable for this component. |
AnalogItemType (DataType=Double) |
Mandatory |
(informative) – Parameter Representation