Annex B (informative) – Example of extending ADI Information Model for gas chromatograph devices ToC Previous Next

B.2 Gas Chromatograph Parameters ToC Previous Next

B.2.2 Parameters defined for a AnalyserChannel of ChromatographDeviceType ToC Previous Next

The example set of Parameters defined on an AnalyserChannel of ChromatographDeviceType for a gas chromatograph are described in Table 109.

Table 109 - ChromatographDeviceType AnalyserChannel Configuration Parameters

BrowseName Description VariableType Optional/
Mandatory
RunState Sets the state of the chromatographic application.
0 = This application is in HOLD
1 = This application is in the RUN state
2 = This application is in the CALibration state
3 = This application is in the VALidation state
Other values are not allowed.
Write format: range 0 - 3
0 = Sets application to HOLD state
1 = Sets application to RUN state
2 = Sets application to CAL state
3 = Sets application to VAL state
DataItemType
(DataType=Int32)
O