Annex A (informative) – Example of extending ADI Information Model for particle size monitor devices

A.2 Parameters of ParticleSizeMonitorDeviceType

A.2.2 AnalyserChannel of ParticleSizeMonitorDeviceType (General Approach)

As an alternative to chapter A.2.1, the tables below show a more general approach of defining the Parameters, independent of the used technology.
Table 96 – ParticleSizeMonitorDeviceType AnalyserChannel Status Parameters (Alternative to Table 94)
BrowseName | Description | VariableType | RW | Optional/ Mandatory |
---|---|---|---|---|
InstrumentConnected | Return the status of the physical connection with the channel | TwoStateDiscreteType (DataType=Boolean) |
RO | O |
ReadyForBackground | Return the status of the instrument and accessories to perform a background reading | TwoStateDiscreteType (DataType=Boolean) |
RO | O |
ReadyForMeasurement | Return the status of the instrument and accessories to perform a measurement | TwoStateDiscreteType (DataType=Boolean) |
RO | O |
All Parameters organized by the Status FunctionalGroup on an AnalyserChannel of a ParticleSizeMonitorDeviceType shall be read-only.
Table 97 – ParticleSizeMonitorDeviceStreamType AcquisitionSettings Parameters (Alternative to Table 95)
BrowseName | Description | VariableType | Optional/ Mandatory |
---|---|---|---|
AcquisitionSettings | Name of a set of acquisition settings stored in the analyser. | String | M |